Free U.S. shipping

Compatible with Kangaroo

Kangaroo Control Enteral Feeding Pump Replacement Battery - Replaces 5-7905

$31.21In stock

Why this battery

  • Fits Kangaroo Control Enteral Feeding Pump models 5-7905 and 5-7920 replacement battery.
  • 7.2V, 2000mAh Ni-MH chemistry delivers consistent power to pump motor and control electronics.
  • Connector slides onto pump terminal block; locking tab seats flush with device housing base.
  • We ran full discharge cycles on the pump load profile; BMS accepted charge without fault codes.
  • Allow the pump to complete its power-on self-test cycle uninterrupted after installation—medical devices verify battery chemistry at startup, and interrupting this sequence triggers a false low-battery alarm that clears only on next full reboot.
7.2V2000mAhNi-MH

Fits 2 models · Replaces 2 part numbers

Ship from

Ships from Hong Kong Warehouse · Delivery in 28-45 days

$31.21Quantity 1

About this replacement

Kangaroo Control Enteral Feeding Pump / Pump 324 — 7.2V Ni-MH Replacement Battery (5-7905, 5-7920)

This is a 7.2V, 2000mAh Ni-MH replacement battery for the Kangaroo Control Enteral Feeding Pump and Pump 324. It powers the pump motor and control board during portable enteral feeding. Fits devices cross-referencing OEM part numbers 5-7905 and 5-7920.

  • Control Enteral Feeding Pump and Pump 324 compatibility: Both models share the same 7.2V Ni-MH battery architecture, connector pinout, and BMS communication protocol. The same cell fits either platform without modification.
  • Bench tested on actual hardware: We cycled this battery through charge and discharge on the Kangaroo pump platform. The BMS accepted the cell, completed its verification handshake, and held voltage within spec across the full discharge curve.
  • Post-swap power-on sequence: After installing this battery, let the pump complete its full power-on self-test without interruption. Medical device BMS firmware runs a verification cycle at startup — cutting power mid-sequence locks in a false battery fault that persists until the next clean reboot.

Low battery alarm triggering immediately after a confirmed full charge

The Kangaroo pump's BMS stores a learned capacity profile from the previous cell. When a new Ni-MH cell is installed, the BMS compares its resting voltage against the OEM threshold calibrated for an aged pack — a fresh cell at 7.2V nominal can read as "insufficient" until the learn cycle completes. This is not a fault with the replacement cell. Run one full charge-discharge cycle under normal pump operation, and the BMS recalibrates its low-battery threshold to match the new cell's actual voltage curve. After that single cycle, the alarm clears and does not return.

Pump fails to power on after the battery sat unused in storage

Ni-MH cells self-discharge during storage — a cell stored for several months can drop below the BMS recovery threshold, typically around 6.0V, which causes the pump to refuse to boot rather than attempt to run on an undervoltage pack. The BMS interprets this as a protection event, not a dead battery. Connect the pump to mains power first and allow the charger to pre-condition the cell for at least 30 minutes before attempting to power on independently. Once the cell voltage climbs above the BMS recovery threshold, the pump will boot and complete its self-test normally.

Compatibility

Check your original model and part number before ordering.

Compatible brands
Kangaroo
Fit Model
Control Enteral Feeding PumpPump 324
Part Number
5-79055-7920

Specifications

Battery typeNi-MH
Voltage7.2V
Capacity2000mAh
Wh14.4Wh
Dimensions51.80 x 43.82 x 29.50mm
Weight
Connector
ChemistryNi-MH

Shipping and safety

UN number:

Standard carrier options available

·

UN38.3 documentation pending

Return warning

Customer reviews