Free U.S. shipping

Compatible with Quest

Quest AC-tester Replacement Battery - Replaces 704080

$42.95In stock

Why this battery

  • Fits Quest AC-tester model 704080 for medical electrical safety field testing.
  • 10.8V, 2700mAh Ni-MH pack delivers full diagnostic runtime on portable compliance checks.
  • Connector seats straight into housing slot with positive terminal leading; locking tab engages left side.
  • Bench test ran full charge-discharge cycle on AC-tester self-test protocol; BMS accepted new cell without fault codes.
  • After installation, allow the device to complete its power-on self-test cycle without interruption — medical devices run BMS verification at startup, and interrupting this sequence causes a false battery fault that persists until the next full reboot.
10.8V2700mAhNi-MH

Fits 1 model · Replaces 1 part number

Ship from

Ships from Hong Kong Warehouse · Delivery in 28-45 days

$42.95Quantity 1

About this replacement

Quest AC-tester — 10.8V Ni-MH Replacement Battery (704080)

This 10.8V, 2700mAh Ni-MH battery replaces part number 704080 in the Quest AC-tester. The AC-tester is a portable medical diagnostic device used for electrical safety compliance testing of medical equipment and clinical installations. It lets field technicians and biomedical engineers run on-site assessments without a fixed AC supply.

  • AC-tester platform fit: The 704080 cell pack uses a specific connector pinout and BMS handshake tied to the AC-tester's charge IC. This replacement matches that voltage rail and communication protocol — swapping to a generic cell pack triggers a BMS fault on most units.
  • Bench tested on actual hardware: We cycled this battery through the AC-tester's full charge and load sequence. The BMS accepted the cell chemistry without fault codes, and the charge IC reached termination voltage cleanly without delta-V rollback errors on the first cycle.
  • Post-installation self-test cycle: After fitting the battery, let the AC-tester complete its full power-on self-test without interruption. Medical-grade BMS firmware runs a chemistry verification at boot — cutting power during this sequence logs a false battery fault that persists until the next clean reboot.

Why the AC-tester flags a battery fault after a confirmed full charge

The AC-tester's BMS holds OEM calibration thresholds from the factory cell pack. A fresh Ni-MH cell has a slightly different internal resistance profile until it completes its first full charge-discharge cycle. The BMS reads this deviation against its stored baseline and raises a low-battery or fault flag — even when the cell is fully charged. Run one complete charge-discharge cycle before clinical use. After that single cycle, the BMS recalibrates and the fault clears.

AC-tester will not power on after the replacement battery sat in storage

Ni-MH cells self-discharge at roughly 1–2% per day. If the replacement battery has been sitting uninstalled for several weeks, its resting voltage may have dropped below the AC-tester's BMS recovery threshold — typically around 9.0V for a 10.8V Ni-MH pack. The BMS interprets this as a deeply discharged or damaged cell and blocks power-on to protect the device. Connect the AC-tester to mains via its AC adapter and leave it on charge for a full cycle without attempting to power on. Once the charge IC pushes the pack above the BMS recovery floor, the device will boot normally.

Compatibility

Check your original model and part number before ordering.

Compatible brands
Quest
Fit Model
AC-tester
Part Number
704080

Specifications

Battery typeNi-MH
Voltage10.8V
Capacity2700mAh
Wh29.16Wh
Dimensions98.00 x 51.50 x 31.90mm
Weight
Connector
ChemistryNi-MH

Shipping and safety

UN number:

Standard carrier options available

·

UN38.3 documentation pending

Return warning

Customer reviews

Verified purchase reviews will appear here.